Instruction Guide – Massa M3 Wireless Sensor Evaluation Kit Page of 14
M3 Kit Guide 111017.docx 10-20-2011
3
PRELIMINARY
M3 Application Software
Validating the Tanks Ultrasonic Performance
This section of the Instruction Guide describes how to evaluate the ultrasonic signals being
measured in the tank, to ensure reliable level measurement. The M3 Software allows access to
sensor status information which reports distance to the liquid surface, air temperature, radio signal
strength, battery voltage, and other status information. The M3 Software also will record the signal
magnitudes vs. time of the ringing caused by the ultrasonic transmit pulse and the echoes reflected
from the targets around the sound path within the tank. This feature can be used to analyze and
determine if signals are sufficient in the application and if there are any issues that may exist, such as
unwanted reflections (see the “Understanding the Ultrasonic Waveform” section).
1) If not already running, start the M3 Application Software (Step 2 on the Page 2).
2) The first time the M3 application software is used, the comm port that connects the USB
converter must be selected. From the dropdown menu, go to “View” and select “Program
Settings”. Go to the “Serial Communications Port” menu, select a comm port, and then click
“OK”, then go to “Search for Sensors”. If the Sensor is not found, go back and select another
comm port and repeat. Once the Sensor is found, all of its settings will be uploaded.
3) When the sensor is found, it will be in the “On-Line
Sensor List”, as shown in the figure on the right.
Check the “Shown” box.
4) Point Sensor at a target such as the ceiling. Click the “Data” tab and then click the “Acquire
Data, Don’t Store” button to measure the distance to the target. Other information will also
be displayed, as shown in the figure below. Clicking on “Acquire Data, Store” performs the
same operation, but also assigns an event number that will be stored in the Sensor.